MAMMOMAT Inspiration Recalled by Siemens Medical Solutions USA, Inc Due to There were instances where the operator table was...

Date: October 24, 2025
Company: Siemens Medical Solutions USA, Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

MAMMOMAT Inspiration;

Quantity: 6 units (3 US, 3 OUS)

Why Was This Recalled?

There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

Where Was This Sold?

This product was distributed nationwide across all 50 states.