Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a Recalled by REPLIGEN CORPORATION Due to Potential Elevated Laser Output A disconnected cable or...

Date: December 10, 2025
Company: REPLIGEN CORPORATION
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact REPLIGEN CORPORATION directly.

Affected Products

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Quantity: 33

Why Was This Recalled?

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Where Was This Sold?

U.S.