Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips Tempus Pro Patient Monitor Recalled by Remote Diagnostic Technologies Ltd. Due to Patient monitor measures Intracranial Pressure (ICP) and Bladder...

Date: November 26, 2025
Company: Remote Diagnostic Technologies Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Remote Diagnostic Technologies Ltd. directly.

Affected Products

Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual

Quantity: 7129

Why Was This Recalled?

Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Remote Diagnostic Technologies Ltd.

Remote Diagnostic Technologies Ltd. has 12 total recalls tracked by RecallDetector.