Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Radiography 7300 C Recalled by Philips Medical Systems DMC GmbH Due to Philips has identified that some ceiling suspension telescopic...

Date: October 1, 2025
Company: Philips Medical Systems DMC GmbH
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Medical Systems DMC GmbH directly.

Affected Products

Radiography 7300 C; Model Number: 712037;

Quantity: 284 units

Why Was This Recalled?

Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

Where Was This Sold?

This product was distributed nationwide across all 50 states.