Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Due to Six (6) identified software issues which may result...

Date: December 12, 2025
Company: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. directly.

Affected Products

Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azurion 3 M15. Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12. Model Number: 722227. 4. Azurion 5 M20. Model Numbers: 722228, 722281. 5. Azurion 7 B12. Model Numbers: 722067, 722225. 6. Azurion 7 B20. Model Numbers: 722068, 722226. 7. Azurion 7 M12. Model Numbers: 722078, 722223. 8. Azurion 7 M20. Model Numbers: 722079, 722224, 722282.

Quantity: 10,466 units

Why Was This Recalled?

Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has 154 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report