Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Conductivity Standard Solution Catalog/Model Numbers: 10001 Recalled by NIPRO Technical Services, Inc. Due to due to a defect associated with the conductivity...

Date: December 12, 2025
Company: NIPRO Technical Services, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact NIPRO Technical Services, Inc. directly.

Affected Products

Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters

Quantity: 499 units

Why Was This Recalled?

due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About NIPRO Technical Services, Inc.

NIPRO Technical Services, Inc. has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report