Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MC3 VitalFlow Console Recalled by Medtronic Perfusion Systems Due to As of August 6, 2025, Medtronic has received...

Date: October 3, 2025
Company: Medtronic Perfusion Systems
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Perfusion Systems directly.

Affected Products

MC3 VitalFlow Console, REF 58100; Blood pump of ecmo

Quantity: 167 units

Why Was This Recalled?

As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.

Where Was This Sold?

This product was distributed nationwide across all 50 states.