Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Restore Clinician Programmer Application (CP App) Recalled by Medtronic Neuromodulation Due to Software issue where a Device Reset message displayed...

Date: November 20, 2025
Company: Medtronic Neuromodulation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medtronic Neuromodulation directly.

Affected Products

Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TABLET UK CT900D PROG CT900D CLINICIAN TABLET GLOBAL CT900E PROG CT900E CLINICIAN TABLET GLBL SMSNG CT900F PROG CT900F CLINICIAN TABLET GLBL SMSNG

Quantity: 7667 devices

Why Was This Recalled?

Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medtronic Neuromodulation

Medtronic Neuromodulation has 95 total recalls tracked by RecallDetector.