Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medline Hudson RCI Medium Concentration Oxygen Masks: 1041CE MASK Recalled by Medline Industries, LP Due to Firm received multiple complaints regarding tubing disconnecting from...

Date: November 21, 2025
Company: Medline Industries, LP
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medline Industries, LP directly.

Affected Products

Medline Hudson RCI Medium Concentration Oxygen Masks: 1041CE MASK,OXYGEN,UND CHIN,MED CO OUS ONLY; HUD1035 O2 MASK,MED CONC,SHORT,PEDI,7' TUB SC; HUD1042 O2 MASK,MED CONC,PEDI,7' TUB SC; HUD1930 O2 MASK,MED CONC,ADULT,7' TUB UC; HUDRHO41U MASK,OXYGEN,MEDIUM-CONCENTRATION,7,SC

Quantity: 1,611,081 eaches

Why Was This Recalled?

Firm received multiple complaints regarding tubing disconnecting from oxygen mask. If issue occurs during use, may lead to delay in patient care and patient may experience shortness of breath leading to hypoxia, which may require medical intervention to prevent further respiratory deterioration.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medline Industries, LP

Medline Industries, LP has 182 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report