Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS Recalled by Medline Industries, LP Due to Medline has identified issues related to calibration of...

Date: January 7, 2026
Company: Medline Industries, LP
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medline Industries, LP directly.

Affected Products

Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS860015W; 4) MAJOR LAPAROTOMY CDS, Model Number: CDS860015X; 5) MINOR LAPAROTOMY CDS, Model Number: CDS860016P; 6) MINOR LAPAROTOMY CDS, Model Number: CDS860016Q; 7) MINOR LAPAROTOMY CDS, Model Number: CDS860016R; 8) MINOR LAPAROTOMY CDS, Model Number: CDS860016S; 9) LAP GASTRIC BYPASS CDS, Model Number: CDS860146M; 10) NEW LONDON MINOR CDS, Model Number: CDS982518M; 11) NEW LONDON MINOR CDS, Model Number: CDS982518N; 12) NEW LONDON MINOR CDS, Model Number: CDS982518O; 13) NEW LONDON MINOR CDS, Model Number: CDS982518P; 14) GENERAL LAPAROSCOPY CDS, Model Number: CDS983311J; 15) LAVH CDS, Model Number: CDS983411G; 16) NHP ROBOTIC GENERAL CDS, Model Number: CDS984244J; 17) ROBOTIC-LF, Model Number: CDS984543C; 18) CYSTO CDS, Model Number: CDS984592G; 19) CYSTO CDS, Model Number: CDS984592I; 20) ROBOTIC, Model Number: CDS985270F; 21) GENERAL LAPAROSCOPY CDS, Model Number: CDS985422I; 22) ROBOTIC CDS, Model Number: CDS985429L; 23) LAP CHOLE CDS, Model Number: CDS985558N; 24) LAP CHOLE CDS, Model Number: CDS985558O; 25) ENDO KIT, Model Number: DYKE1580A; 26) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105; 27) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105A; 28) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105B; 29) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116; 30) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116B; 31) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117A; 32) KIT SURG ONC DIAGNOSTIC LAPARO, Model Number: DYKMBNDL12; 33) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121; 34) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121A; 35) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121B; 36) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121C; 37) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121F; 38) KIT L&D CYSTO SETUP, Model Number: DYKMBNDL135A; 39) KIT THORACIC LAPAROSCOPIC HIAT, Model Number: DYKMBNDL136; 40) KIT THORACIC LAPAROSCOPIC HIAT, Model Number: DYKMBNDL136A; 41) KIT THORACIC STAGING LAPAROTOM, Model Number: DYKMBNDL15; 42) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153A; 43) KIT SURG ONC LAP DIAG/PORT INS, Model Number: DYKMBNDL153B; 44) KIT LAP DONOR NEPHRECTOMY, Model Number: DYKMBNDL173; 45) KIT LAP DONOR NEPHRECTOMY, Model Number: DYKMBNDL173A; 46) KIT THORACIC HIATAL HERNIA TRA, Model Number: DYKMBNDL175; 47) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176; 48) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176A; 49) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176B; 50) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176C; 51) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184; 52) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184A; 53) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188A; 54) KIT UROLOGY LAP SIDDIQUI, Model Number: DYKMBNDL193; 55) KIT UROLOGY LAP SIDDIQUI, Model Number: DYKMBNDL193A; 56) KIT UROLOGY BUCCAL GRAFT, Model Number: DYKMBNDL206; 57) KIT STC PLASTIC FREE FLAP, Model Number: DYKMBNDL54A; 58) KIT UROLOGY CYSTOSCOPY SUPPLEM, Model Number: DYKMBNDL60; 59) KIT UROLOGY CYSTOSCOPY SUPPLEM, Model Number: DYKMBNDL60A; 60) KIT UROLOGY MINOR, Model Number: DYKMBNDL61A; 61) KIT UROLOGY MINOR, Model Number: DYKMBNDL61B; 62) KIT GEN SURG LAP COLECTOMY, Model Number: DYKMBNDL73; 63) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73A; 64) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73B; 65) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73C; 66) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73D; 67) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74; 68) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74A; 69) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74B; 70) KIT GEN SURG LAP GASTRIC BYPAS,

Quantity: 82597 units

Why Was This Recalled?

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medline Industries, LP

Medline Industries, LP has 182 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report