Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medline 14 Fr Suction Catheter Kits SKU DYND40972 - respiratory convenience kit Recalled by Medline Industries, LP Due to A smaller open suction catheter, the 8 Fr...

Date: December 24, 2025
Company: Medline Industries, LP
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Medline Industries, LP directly.

Affected Products

Medline 14 Fr Suction Catheter Kits SKU DYND40972 - respiratory convenience kit

Quantity: 141,971

Why Was This Recalled?

A smaller open suction catheter, the 8 Fr Suction Catheter, may have been packed within the kit instead of the required 14 Fr Suction Catheter. Use of the smaller size catheter may result in hypoxia and/or respiratory failure related to partial or full airway obstruction. Airway obstruction may require immediate emergent medical intervention to prevent respiratory failure or death. Other potential outcomes include prolonged procedure, airway irritation, or prolonged dyspnea.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Medline Industries, LP

Medline Industries, LP has 182 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report