Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brand Name: Da Vinci Product Name: da Vinci X Recalled by Intuitive Surgical, Inc. Due to Due to a software implementation error that allowed...

Date: November 7, 2025
Company: Intuitive Surgical, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Intuitive Surgical, Inc. directly.

Affected Products

Brand Name: Da Vinci Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgical System Model/Catalog Number: IS4000; IS4200; IS5000 Software Version: N/A Product Description: The da Vinci Xi, X, and da Vinci 5 surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument

Quantity: 219

Why Was This Recalled?

Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction

Where Was This Sold?

This product was distributed to 30 states: AL, AR, CA, CO, CT, FL, ID, IL, IN, KS, KY, MD, MI, MN, MS, MO, NV, NJ, NY, NC, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI