GEM Premier 5000 PAK Recalled by Instrumentation Laboratory Due to Confirmed customer complaints indicating that GEM PAKs (cartridges)...

Date: January 20, 2026
Company: Instrumentation Laboratory
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Instrumentation Laboratory directly.

Affected Products

GEM Premier 5000 PAK, Part No. 00055360004.

Quantity: 51 units

Why Was This Recalled?

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Where Was This Sold?

This product was distributed nationwide across all 50 states.