Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

FDR Visionary Suite Recalled by FUJIFILM Healthcare Americas Corporation Due to It was found that the contacts of the...

Date: January 9, 2026
Company: FUJIFILM Healthcare Americas Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact FUJIFILM Healthcare Americas Corporation directly.

Affected Products

FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)

Quantity: 16 units

Why Was This Recalled?

It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About FUJIFILM Healthcare Americas Corporation

FUJIFILM Healthcare Americas Corporation has 19 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report