Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: Recalled by ETAC A/S Due to A customer experienced that during assembly and load...

Date: November 17, 2025
Company: ETAC A/S
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ETAC A/S directly.

Affected Products

Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;

Quantity: 95 units

Why Was This Recalled?

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About ETAC A/S

ETAC A/S has 4 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report