Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brand Name: Catalyst+ Product Name: Catalyst Recalled by C-RAD POSITIONING AB Due to Due to issues with the system's stereotactic radiosurgery...

Date: October 9, 2025
Company: C-RAD POSITIONING AB
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact C-RAD POSITIONING AB directly.

Affected Products

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Quantity: 189

Why Was This Recalled?

Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About C-RAD POSITIONING AB

C-RAD POSITIONING AB has 3 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report