Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60 Recalled by Cellavision AB Due to Automated cell-locating device barcode reader may read the...

Date: October 8, 2025
Company: Cellavision AB
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cellavision AB directly.

Affected Products

REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.

Quantity: 688 units

Why Was This Recalled?

Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Cellavision AB

Cellavision AB has 2 total recalls tracked by RecallDetector.

Related Recalls