Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm Recalled by Boston Scientific Corporation Due to Increased reports of stent deployment and expansion issues....

Date: December 19, 2025
Company: Boston Scientific Corporation
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Boston Scientific Corporation directly.

Affected Products

AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;

Quantity: 557 units

Why Was This Recalled?

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Boston Scientific Corporation

Boston Scientific Corporation has 601 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report