Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

COULTER DxH Diluent Recalled by Beckman Coulter, Inc. Due to Due to specific lots of diluents contributing to...

Date: November 10, 2025
Company: Beckman Coulter, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Beckman Coulter, Inc. directly.

Affected Products

COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.

Quantity: 76,098 units (US-61843 and OUS-14255)

Why Was This Recalled?

Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Beckman Coulter, Inc.

Beckman Coulter, Inc. has 79 total recalls tracked by RecallDetector.

Related Recalls