Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brand name: AGFA Digital Radiography X- Ray system DR 800 Recalled by AGFA Healthcare Corp. Due to It is possible to exceed the maximum allowed...

Date: November 18, 2025
Company: AGFA Healthcare Corp.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact AGFA Healthcare Corp. directly.

Affected Products

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Quantity: 35

Why Was This Recalled?

It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.

Where Was This Sold?

This product was distributed nationwide across all 50 states.